Description
The Agilent 1200 Series vacuum degasser, model G1322A, comprises a 4-channel vacuum container, including 4 tubular plastic membranes, and a vacuum pump. When the vacuum degasser is switched on, the control circuit turns on the vacuum pump which generates a partial vacuum in the vacuum container The pressure is meassured by a pressure sensor. The vacuum degasser maintains the partial vacuum by turning on and off the vacuum pump depending on the signal from the pressure sensor.
Agilent G1322A Degasser
The LC pump draws the solvents from their bottles through the special tubular plastic membranes of the vacuum container. As the solvents pass through the vacuum tubes any dissolved gas in the solvents permeates through the membranes into the vacuum container. The solvents will be almost completely degassed when leaving the outlets of the vacuum degasser.
G1322A Degasser
Performance specifications
Maximum flow rate | 10 ml/min per channel |
Number of channels | 4 |
Internal volume per channel | Typically 12 ml per channel |
Materials in contact with solvent | PTFE, PEEK |
pH range | 1 – 14 |
Analog output (AUX) | For pressure monitoring, range 0 – 3 V |
Physical specifications
Weight | 7 kg | |
Dimensions | 345 × 435 × 80mm | |
Line Voltage | 100–120 or 220– 240 VAC, ± 10% | Wide-ranging capability |
Line Frequency | 50 or 60Hz,±5% | |
Power consumption | 30W | Maximum |
Ambient Operating Temperature | 0–55°C(32–131°F)1 | |
Ambient Non-operating Temperature | -40–70°C(-4–158°F) | |
Humidity | <95%,at25–40°C(77–104°F) | Non-condensing |
Operating Altitude | Up to 2000 m (6500 ft) | |
Non-operating Altitude | Up to 4600 m (14950 ft) | For storing the instrument |