Hoofdafbeelding Agilent G1362A RID HPLC 1200 series

Agilent G1362A RID HPLC 1200 series

Article nr.: 8035
Price on request
Accurate
Flexible setup for different analyzes
Software friendly

Description

Agilent 1200 Series G1362  Refractive Index Detector (RID) The Agilent 1200 Series Refractive Index Detector (RID) is the ideal detector for fast and reliable LC results when routinely analyzing non-UV absorbing substances, such as carbohydrates, lipids, and polymers. The 1200 Series RID is also the detector of choice for gel permeation chromatography (GPC) or size exclusion chromatography (SEC).

Agilent G1362A RID 

Refractive index range1.00 – 1.75 RIU, calibrated
Measurement range+/- 600 x 10-6 RIU
Optics temperture control5 °C above ambient to 55 °C
Sample cellVolume 8 µL, maximum pressure 5 bar (0.5 Mpa), maximum flow rate 5 mL/min
ValvesAutomatic purge and automatic
VolumesInlet port to sample cell 62 µL / inlet port to outlet port 590 µL 
Liquid contact materials316 stainless steel, PTFE and quartz glass
pH range2.3 – 9.5
Performance specifications

Short term noise < +/- 2.5 x 10-9 RIU

Drift < 200 x 10-9 RIU/h

Time programmable parameters

Polarity, peak width

Detector zeroAutomatic zero before analysis
Control and data evaluationParameter entry, signal display, on-line help and diagnostics with the Agilent 1260 Infinity
Control Module. Optional PCMCIA card for method, equence and logbook storage
and transfer. Agilent ChemStation for LC PC based software for control and data
evaluation.
Analog outputsRecorder/integrator: 100 mV or 1 V, output range selectable, one output
CommunicationsController-area network (CAN), LAN, RS-232C, APG Remote: ready, start, stop and shut-down signals
Safety and maintenance Extensive diagnostics, error detection and display (through control module and
ChemStation), leak detection, safe leak handling, leak output, signal for shutdown of pumping
system. Low voltages in major maintenance areas.
GLP featuresEarly maintenance feedback (EMF) for continuous tracking of instrument usage with
user-selectable limits and feedback messages. Electronic records of maintenance and errors. Automated operational qualification/performance verification (OQ/PV).

 

Want to stay up to date with the latest news?
Then sign up for our newsletter!
Stay informed and sign up for our newsletter