G1379B Degasser

The Agilent 1200 HPLC G1379B Micro Vacuum Degasser comprises a 4-channel vacuum container, including 4 micro structured membranes, and a vacuum pump. When the micro vacuum degasser is switched on, the control circuit turns on the vacuum pump, which is operated with constant speed and voltage, generating a partial vacuum in the vacuum container. The degasser maintains the partial vacuum with the help of a controlled leak in the degasser's proportional valve. 

Performance specifications G1379B Degasser

Max Flow Rate0-5 ml/min per channel
Number of Channels4
Internal Volume per ChannelTypically 1 ml per channel
Materials in Contact with SolventPTFE, FEP, PEEK
pH Range1 – 14
RS-232 OutputFor diagnostic purposes
Dimensions (W x D x H)13.5 x 17 x 3.1 in (345 x 435 x 80 mm)
Electrical Requirements100-120 or 220-240 VAC, 50/60 Hz

* Click for more information on the Agilent G1379B Degasser

Specifications G1312B Binary Pump

Hydraulic systemTwo dual piston in series pumps with proprietary servo-controlled variable stroke drive, floating piston design and active inlet valve
Setable flow rangeSetpoints 0.001 – 5 ml/min, in 0.001 ml/min increments
Flow range0.1 – 5.0 ml/min
Flow precision£ 0.07% RSD, or £ 0.02 min SD whatever is greater, based on retention time at constant room temperature
Flow accuracy± 1% or 10 μl/min whatever is greater
PressureOperating range 0 400 bar (0 – 5880 psi) up to 5 ml/min
Pressure pulsation< 2 % amplitude (typically < 1 %), at 1 ml/min isopropanol, at all pressures > 1 MPa
Compressibility compensationUser-selectable, based on mobile phase compressibility
Recommended pH range1.0 – 12.5, solvents with pH < 2.3 should not contain acids which attack stainless steel
Gradient formationHigh-pressure binary mixing, delay volume 180 – 480 μl without mixer, 600 – 900 μl with mixer, dependent on back pressure
Composition range1 – 99 % or 5 μl/min per channel, whatever is greater
Composition precision£ 0.5% absolute
Composition accuracy± 0.15% RSD, at 1 ml/min
Control and data evaluationAgilent ChemStation for LC
Analog outputFor pressure monitoring, 2 mV/bar, one output
Communications

Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready, start, stop and shut-down signals, LAN optional